The Trilogy 202 is both a volume-control & pressure-control ventilator for invasive and noninvasive ventilation. Versatile breath delivery and setup options provide greater continuity of care.
CPAP | 4 - 20 cmH20 (passive leak port circuits)
cm H₂O
|
IPAP | 50
cm H₂O
|
EPAP/PEEP (a) | 0 - 25 cmH20 (active valve circuits)
cm H₂O
|
EPAP/PEEP (b) | 4 - 25 cmH20 (passive leak port circuits)
cm H₂O
|
Pressure support | 0 – 30 cmH2O
cm H₂O
|
Tidal volume | 50 - 2000
ml
|
Breath rate(a) | 0 - 60 BPM (AC mode)
beats per minute
|
Breath rate(b) | 1 - 60 BPM (all other modes)
beats per minute
|
Inspiratory time | 0.3 - 5.0 s Synchrony features
|
Rise time | 1 - 6 (relative scale)
|
Ramp start pressure (a) | 0 - 25 cmH20 (active circuits)
cm H₂O
|
Ramp start pressure (b) | 4 - 25 cmH20 (passive circuits)
cm H₂O
|
Ramp start pressure (c) | 4 - 19 cmH20 (CPAP mode)
cm H₂O
|
Ramp time Off | 5 - 45
min
|
C-Flex Off | 1 - 3 (relative scale)
|
Flow trigger sensitivity | 1 - 9
l/min
|
Flow cycle sensitivity | 10 - 90
%
|
Operating temperature | 5 – 40
°C
|
Relative humidity | 15 - 95
%
|
Storage temperature | -20 – 60
°C
|
Atmospheric pressure | 60 - 110 kPa (450 - 825 mmHg)
|
Adult | Yes
|
Pediatric | ≥ 5 kg or greater
|
Input voltage | 100 - 240 VAC and 50/60 Hz and 2.1 A
|
Detachable battery voltage | 14.4 VDC
|
Internal battery life | 3 hours under normal conditions
hr
|
Detachable battery life | 3 hours under normal conditions
hr
|
external battery connection | 12 VDC
|
Pressure control (PC) | Yes
|
MPV Mouthpiece Ventilation Mode | Yes
|
Synchronized intermittent mandatory ventilation (SIMV) | Yes
|
SIMV with pressure support (SIMV w/PS) | Yes
|
Control ventilation (CV) | Yes
|
Low tidal volume Off | 50 - 2000
ml
|
Circuit disconnect Off | 10 - 60
sec
|
Apnea Off | 10 - 60 s and 4 - 60 BPM
|
High tidal volume Off | 50 - 2000
ml
|
High minute ventilation Off | 1 - 99 l/min Off, 1 - 99
l/min
|
Low minute ventilation Off | 1 - 99
l/min
|
High respiratory rate Off | 4 - 80
beats per minute
|
Low respiratory rate Off | 4 - 80
beats per minute
|
Pressure control (PC) | Yes
|
Pressure control-SIMV (PC-SIMV) | Yes
|
Spontaneous ventilation (S) | Yes
|
Spontaneous ventilation with timed back-up (S/T) | Yes
|
Timed ventilation (T) | Yes
|
Continuous positive airway pressure (CPAP) | Yes
|
Average volume assured pressure support (AVAPS) | With passive circuit type, S, S/T, PC, and T modes only
|
Detachable back-up battery | Offering up to 3 hours additional operating time
|
Hospital roll stand | Provides convenient accessory basket and humidifier mount
|
DirectView | 1 GB SD card data storage integrated into the ventilator software
|
Tidal volume | 0 - 2000
ml
|
Minute ventilation | 0 - 99
l/min
|
Estimated leak rate | 0 - 200
l/min
|
Respiratory rate | 0 - 80 BPM 0 - 80
beats per minute
|
Peak inspiratory flow | 0 - 200
l/min
|
Peak inspiratory pressure | 0 - 99
cm H₂O
|
Mean airway pressure | 0 - 99
cm H₂O
|
% patient triggered breaths | 0 - 100
%
|
I:E ratio | 9.9:1 - 1:9.9
|
Auto-Trak sensitivity | Auto-Trak sensitivity Auto-adaptive triggering, cycling, and leak compensation (available in all modes, passive circuit only)
|
Adjustable flow triggering | 1 – 9 l/min (available in all modes and circuit types)
l/min
|
Active exhalation valve with proximal airway pressure (PAP) | Yes
|
Active exhalation valve with flow sensor | Yes
|
Passive exhalation port | Yes
|
IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety | Yes
|
IEC 60601-1-2 General requirements for safety – collateral standard Electromagnetic compatibility – requirements and tests | Yes
|
IEC 60601-2-12 Medical electrical equipment Part 2-12: Particular requirements for the safety of lung ventilators – Critical Care ventilators | Yes
|
FiO2 | 21 - 100
%
|
02 flush | 2 min at 100%
|
02 input pressure range | 276 - 600 kPa (40 - 87 psi)
|