Thuisbeademing

Trilogy 200

Portable ventilator

Product nummer:

1040005

Versatile, easy-to-use Trilogy 200 provides invasive and non-invasive ventilator support for adult and pediatric patients in the home, hospital and acute-care sites. The pressure- and volume-mode ventilator has proprietary therapy algorithms, connectivity and enhanced triggering sensitivity which decreases the work of breathing. Trilogy 200 helps care teams provide quality respiratory care while meeting patients’ evolving needs.

Contactinformatie

* Dit veld is verplicht
*

Contactgegevens

*
*
*

Bedrijfsgegevens

*
*
*
*
*

Informatie details

Door uw reden voor contact op te geven, kunnen wij u beter van dienst zijn.
*
*
Kenmerken
Reliable and portable
Continuous respiratory support for added freedom and mobility

Continuous respiratory support for added freedom and mobility

Trilogy provides respiratory support to meet your prescribed needs, up to 24 hours a day. Trilogy is portable with internal, hot-swappable detachable, and external batteries to provide greater freedom and mobility with continuous ventilation. The easy-to-use carry bag is wheelchair-mountable and designed so the screen is always visible.
Volume Targeting
Adapting with every breath

Adapting with every breath

AVAPS-AE--an auto-titration mode of non-invasive ventilation with multiple algorithms that work simultaneously--meets the changing needs of your patients. AVAPS automatically adjusts to breathing, while Auto EPAP proactively adjusts to the lowest effective pressure for peak comfort. Auto Backup delays a machine breath until the patient has exhaled to reduce air trapping.
Flexible
Easy to use and adaptable for your patients’ changing needs

Easy to use and adaptable for your patients’ changing needs

Trilogy is designed with easy-to-read and navigate screens, multiple Rx settings and on-screen waveforms to help optimize treatment. Trilogy’s pressure- and volume-control features allow you to provide more types of therapy with one device. The ability to use multiple modes for patients who require varying levels of ventilation, allows greater flexibility in care.
Mouthpiece ventilation
Delivering on-demand ventilation

Delivering on-demand ventilation

With no inspiratory effort required to trigger a breath, our unique ‘kiss’ trigger detects when a patient engages and disengages from the mouthpiece to deliver on-demand ventilation with the security of patient alarms.
Digital Auto-Trak
When sensitivity is needed most

When sensitivity is needed most

This adaptive algorithm automatically adjusts the triggering and cycling of the ventilator, while compensating for leaks. Digital Auto-Trak’s capability contributes to the patient’s synchrony and comfort, without requiring manual adjustments. The Sensitive Auto-Trak setting provides an enhanced triggering response for patients with minimal respiratory effort.
Supported
The support and training you need

The support and training you need

The innovative technology of the Trilogy family sets it apart from the rest. The support, service and upkeep behind all of it keeps it there, including 24/7 clinical support, educational training resources for both you and your patients and extended warranties and maintenance plans.
Accessories
Enhance patient mobility and care with Trilogy accessories

Enhance patient mobility and care with Trilogy accessories

From power options and circuit adapters, to reporting software and alarms, Trilogy’s array of accessories helps give patients and families more freedom and mobility while maintaining effective treatment.

Specifications Bekijk meer

Dimensions and weight
Dimensions
16.68 cm L x 28.45 cm W x 23.52 cm H (6.6" L x 11.2" W x 9.5" H)
Weight
5 kg (11 lb)
Controls
Circuit types
Active Flow; Active with PAP; Passive
Ventilation types
Pressure and Volume
Volume modes
AC; SIMV (w/PS); CV; AC-MPV
Pressure modes
CPAP; S; S/T; T; PC-SIMV (w/PS); PC; PC-MPV; AVAPS-AE
IPAP
4 to 50 cm H₂O
PEEP/EPAP for active circuits
0 to 25 cm H₂O
PEEP/EPAP for passive circuits
4 to 25 cm H₂O
CPAP for passive circuits
4 to 20 cm H₂O
Pressure support differential
0 to 40 cm H₂O
Alarms
Circuit disconnect
Off; 5 to 60 sec
Apnea rate
Off; 10 to 60 sec
High tidal volume
Off; 50 to 2000 ml
Low tidal volume
Off; 40 to 2000 ml
High minute ventilation
Off; 1 to 99 l/min
Low minute ventilations
Off; 0.2 to 99 l/min
High respiratory rate
Off; 4 to 80 BPM
Low respiratory rate
Off; 4 to 80 BPM
Measured patient parameters
Respiratory rate
0 to 80 BPM
Minute ventilation
0 to 99 l/min
Vte/Vti
0 to 2000 ml
Leak rate
0 to 200 l/min
Peak inspiratory flow
0 to 200 l/min
Peak inspiratory pressure
0 to 99 cm H₂O
Mean airway pressure
0 to 99 cm H₂O
% Patient triggered breaths
0 to 100 %
I:E ratio
9.9:1 to 1:9.9
Controls
Tidal volume
0 to 2000 ml
Breath rate
0 to 60 BPM for AC mode; 1 to 60 BPM for all other modes
Inspiratory time
0.3 to 5.0 sec
Rise time
1 to 6
Ramp start pressure
0 to 25 cm H₂O for active circuits; 4 to 25 cm H₂O for passive circuits; 4 to 19 cm H₂O in CPAP
Ramp length
Off; 5 to 45 min
Flex
Off; 1 to 3
Flow trigger sensitivity
1 to 9 l/min
Flow cycle
10 to 90 %
Apnea rate
4 to 60 BPM